Using the spectroelectrochemical flow cell, it is possible to have a different optical path length changing the gasket. We offer, as an optional item, a silicon and Teflon gasket with a 100, 250 and 500 µm of the thickness. SEC-2F Spectroelectrochemical flow cell was designed to fit perfectly in the SEC2000 Spectrometer, and it eliminated the use of the optical fiber to connect the Spectrometer to the SEC-2F. Even for another brand of the spectrometer, you can connect the SEC-2F using the collimating lens and optical fiber.
Catalog No. | Description | |
012660 | SEC-2F Spectroelectrochemical flow cell | |
Contensts | Quantity | |
- | SEC-2F flow cell block A | 1 pc |
- | SEC-2F flow cell block B | 1 pc |
- | SEC-2F flow cell base | 1 pc |
- | SEC-2F flow cell cover | 1 pc |
- | SEC-2F S500 Silicone Gasket | 2 pcs |
- | Stainless tube OD 1/16" (50 mm) | 1 pc |
- | Needle adaptor | 1 pc |
- | Fitting PEEK | 2 pcs |
- | Silicon tube (30 cm) | 1 pc |
- | Teflon tube (1 m) | 1 pc |
For the system set-up, spectrometer (detector), light source, optical fiber, electrochemical analyzer, spectroelectrochemical flow cell, the liquid feed pump (or syringe), and the PC for data processing are required. ITO electrode or platinum, gold or carbon grid electrode could be used as a working electrode.
Grid type electrode and ITO electrode are using in the spectroelectrochemical flow cell.
The light transmittance was compared with ITO electrode and Grid electrode (Au, Pt and Carbon) on a quartz glass as a reference.
For ITO electrode the light cannot be transmitted easily in an ultraviolet range. The transmittance is about 10% at 280 nm of the wavelength. For wavelength above 400 nm the transmittance is above 80%.
Compared with the quartz glass, the light transmission for grid electrode is about 30%, however it could be used in an ultraviolet range.
Measurement of the absorbance for potassium ferrocyanide electrolysis was performed using
100, 250 and 500 µm thickness gasket.
At the left diagram, the characteristic absorption peak of the sample is enclosed at a red line. At the right diagram, the equilimbrium time for electrolysis is shown.
Electrolysis of Vitamin B12 derived complex.